BD MAX Enteric Bacterial Panel (2019-01-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
8199682
Manufacturer
Geneohm Sciences Canada, Inc.,
2555 Boul. Du Parc Technologique,
Québec, Québec
G1P 4S5
CANADA
BD MAX Enteric Bacterial Panel442963

Introduction

Affected Products

BD MAX Enteric Bacterial Panel

Reason

BD has observed a shift in PCR metrics for BD MAX Extended Enteric Bacterial Panel master mixes when used with sample buffer tubes from the affected BD MAX Enteric Bacterial Panel (EBP) lot. The result of this shift may cause false negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides targets of the BD MAX Extended Enteric Bacterial Panel assay.