Health Products

neoBLUE blanket LED Phototherapy System (2019-01-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Natus Medical Incorporated
5900 First Avenue South
Seattle, Washington
98108
UNITED STATES
neoBLUE blanket LED Phototherapy System006244

Introduction

Affected Products

neoBLUE blanket LED Phototherapy System

Reason

This notification is related to a prior notification, October 2016, regarding the neoBLUE blanket LED Phototherapy System.  At that time, Natus Medical informed you of a potential issue of discoloration/degradation/melting of the fibre optic bundle at the pad connector which is inserted in the neoBLUE blanket light box.  Natus Medical completed a redesign of the device and is initiating this field action to replace those systems shipped from 2011 through march 21, 2016.  Newer systems shipped after March 21, 2016 do not need to be replaced.  Natus will provide an updated literature kit for each of the newer devices.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime