Medtronic Programmers (CareLink, Encore), Medtronic Micra (2019-01-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis, Minnesota
55432
UNITED STATES
B. MicraMC1VR01
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis, Minnesota
55432
UNITED STATES
C. CareLink Encore Programmer29901A
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis, Minnesota
55432
UNITED STATES
A. CareLink Programmer2090

Introduction

Affected Products
  1. CareLink Programmer
  2. Micra
  3. CareLink Encore Programmer

Reason

Notification to inform customers of the availability of a software update for Medtronic 2090 CareLink and 29901 Encore Programmers. The update is specific to the programmer software that supports the Micra Transcatheter Pacing System (TPS) device and is released under the application software ID: Model SW022 version 8.1. This software update corrects an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its recommended replacement time (RRT). A software update is required to ensure that the correct remaining longevity estimate is displayed at all times during the life of the device. Without the software update, the displayed value during this period may remain static and reflect an over-estimation of the remaining longevity. Once the RRT voltage threshold is reached the correct remaining longevity estimate is once again displayed. The software display error does not reflect any issues with the Micra TPS device battery, its ability to deliver therapy and to function as designed. The observed performance of the Micra TPS remains within published expectations.