Health Products

LIFEPAK 15 Monitor/Defibrillator (2019-01-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Physio-Control Inc.
11811 Willows Road North East
Redmond, Washington
98052
UNITED STATES
LIFEPAK 15 Monitor/Defibrillator99577-000837

Introduction

Affected Products

LIFEPAK 15 Monitor/Defibrillator

Reason

Stryker has become aware [via customer complaint] that certain LIFEPAK 15 Monitor/Defibrillators were reported to experience a lock-up condition after a defibrillation shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power to the device, but no response in the keypad and device functions. A device in this condition has the potential to delay delivery of therapy, and this delay in therapy has the potential to result in serious injury or death.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime