Health Products

VITEK 2 Test Cards (2019-01-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2440811203
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham, North Carolina
27712
UNITED STATES
D. VITEK 2 ANC ID21347
2410602403
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham, North Carolina
27712
UNITED STATES
E. VITEK 2 GN Card21341
2750809103
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham, North Carolina
27712
UNITED STATES
B.  VITEK 2 Compact 60 System - susceptibility test cards415 670
2880809203
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham, North Carolina
27712
UNITED STATES
C. VITEK 2 System - susceptibility test cards420 739
6260813203
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham, North Carolina
27712
UNITED STATES
A. VITEK 2 Compact 30 System - susceptibility test cards413 143

Introduction

Affected Products
  • A. VITEK 2 Compact 30 System - susceptibility test cards
  • B. VITEK 2 Compact 60 System - susceptibility test cards
  • C. VITEK 2 System - susceptibility test cards
  • D. VITEK 2 ANC ID
  • E. VITEK 2 GN Card

Reason

A potential issue was identified related to the top seal of card pouches containing the VITEK 2 test cards. BioMérieux has determined that the integrity of some of the VITEK 2 test card top seal may be compromised.  A compromised test card top seal can impact card reagents due to the entry of moisture. BioMérieux investigation concluded that the pouch top seal of the card lots manufactured on a specific card poucher (poucher 1) between 10 Aug. 2018 and 29 Aug. 2018 is potentially compromised.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime