Plain Gut Absorbable Collagen Suture (2019-02-12) Feb 12, 2019 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number 170927-52 Manufacturer CP Medical Inc. 1775 Corporate Drive, Suite 150 Norcross 30093 Georgia UNITED STATES Plain Gut Absorbable Collagen Suture 810PG
Affected Products
Plain Gut Absorbable Collagen Suture
Routine monitoring of plain gut sutures not released to market exceeded endotoxin limits. As a precaution, suture lots manufactured from the same raw material lot are being recalled.