Luminex recently identified a manufacturing defect. An internal investigation identified a defect in the back seal that affects a portion of cassettes manufactured between 9/18/2018 and 1/15/2019. The back seal plays a role in maintaining performance until the labeled expiration date. The investigation has concluded performance of the lots listed meet performance claims at this time. However, it is possible that the seal performance may not be maintained throughout the shelf life provided in the product labeling. The identified defect has a remote possibility of leading to an invalid or false negative result when using the affected aries assay lots.