xTAG Gastrointestinal Pathogen Panel (GPP) (2019-02-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
IK032C-2031
Manufacturer
Luminex Molecular Diagnostics, Inc.
439 University Avenue
Toronto
M5G 1Y8
Ontario
CANADA
xTAG Gastrointestinal Pathogen Panel (GPP)I032C0415

Introduction

Affected Products

xTAG Gastrointestinal Pathogen Panel (GPP)

Reason

Luminex received customer complaints regarding xTAG Gastrointestinal Pathogen Panel (GPP) reporting lower MS2 MFI values. Investigational testing has determined that this is due to the variability of conductivity found in a lot of xTAG GPP Reporter Buffer, a component of the xTAG Gastrointestinal Pathogen Panel (GPP) kit. The potential impact of xTAG GPP Reporter Buffer with low conductivity on the xTAG GPP assay is that when testing patient specimens near the limit of detection (LoD) the assay has a remote possibility to generate false negative calls.