Health Products

UniCel DxI 600, 800 Access Immunoassay System - Thyroglobulin Antibody II (2019-02-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
831710
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea, California
92821
UNITED STATES
B. Access System - Thyroglobulin Antibody IIA32898
831710
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea, California
92821
UNITED STATES
C. UniCel DxI 600 Access Immunoassay System - Thyroglobulin Antibody IIA32898
831710
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea, California
92821
UNITED STATES
A. UniCel DxI 800 Access Immunoassay System - Thyroglobulin Antibody IIA32898

Introduction

Affected Products
  1. UniCel DxI 800 Access  Immunoassay System - Thyroglobulin Antibody II
  2. Access System - Thyroglobulin Antibody II
  3. UniCel DxI 600 Access  Immunoassay System - Thyroglobulin Antibody II

Reason

Beckman Coulter has determined through customer feedback and an internal investigation that the four affected access thyroglobulin antibody II (TgAbII) reagent lots may generate falsely decreased results in some patient samples when compared to previously released reagent lots.

The affected lots of TgAbII reagent have been shown to generate repeatable (falsely decreased) results by up to 95% in some patient samples.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime