VITROS 3600 Immunodiagnostic System, VITROS 5600 Integrated System, VITROS XT 7600 Integrated System (2019-02-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial# 56003270 and higher.
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester, New York
14626
UNITED STATES
B. VITROS 5600 Integrated System680 2413
Serial# 76000108 and higher.
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester, New York
14626
UNITED STATES
C. VITROS XT 7600 Integrated System684 4461
Serial# 36001160 and higher.
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester, New York
14626
UNITED STATES
A. VITROS 3600 Immunodiagnostic System680 2783

Introduction

Affected Products
  1. VITROS 3600 Immunodiagnostic System
  2. VITROS 5600 Integrated System
  3. VITROS XT 7600 Integrated System

Reason

A luminometer malfunction may occur after the system is shut down and restarted, due to the luminometer signal board parameters incorrectly reverting to the default settings. If this occurs, users will not be able to process any microwell assays until service is performed. In this scenario, an ortho-trained service representative must be dispatched to resolve the malfunction.