CELL-DYN Emerald System - Instrument (2019-02-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All serial numbers below 7765.
Manufacturer
Abbott Laboratories Diagnostics Division
100 ABBOTT PARK ROAD
ABBOTT PARK
60064
Illinois
UNITED STATES
CELL-DYN Emerald System - Instrument09H39-01

Introduction

Affected Products

CELL-DYN Emerald System - Instrument

Reason

Abbott Hematology has identified occurrences where the CELL-DYN Emerald analyzer generates quality control (QC) low or out of range low for parameters RBC and PLT.

Abbott has identified that preventative maintenance specified in the CELL-DYN Emerald operator's manual and associated cleaning methods may not be sufficient for some of the CELL-DYN Emerald users, to maintain the CELL-DYN Emerald analyzer operational on a routine basis.

This issue does not impact patient results; however, it can cause a delay in the generation of patient results.