Health Products

QUANTA Flash DFS70 Controls (2018-11-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
171589QUANTA Flash DFS70 Controls
Manufacturer
Inova Diagnostics Inc.
9900 Old Grove Road
San Diego
92131
California
UNITED STATES
701267

Introduction

Affected Products

QUANTA Flash DFS70 Controls

Reason

Upon investigation into complaints from customers, Inova Diagnostics has confirmed that the positive control value on lots 171589, 181676 and 181799 of QUANTA Flash DFS70 Controls fall below the acceptable range. When control values fall outside of the acceptable range, the run is considered invalid. Results from an invalid run cannot be reported and the run needs to be repeated.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime