Health Products

MultiDiagnost-ELEVA, OmniDiagnost Systems (2019-03-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
D. MultiDiagnost-ELEVA - SystemMRC 3011
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
E. OmniDiagnost ELEVA708027
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
B. MultiDiagnost-ELEVA Intensifier - TV Image Dectector0708MD02
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
C. MultiDiagnost-ELEVA - Flat Detector - Image Detector0708MD02
Not applicable.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. Philips OmniDiagnostOMNIDIAGNOST

Introduction

Affected Products
  1. Philips OmniDiagnost
  2. MultiDiagnost-ELEVA Intensifier - TV Image Dectector
  3. MultiDiagnost-ELEVA - Flat Detector - Image Detector
  4. MultiDiagnost-ELEVA - System
  5. OmniDiagnost ELEVA

Reason

For the height and tilt movement of the stand/table, the MultiDiagnost-ELEVA, OroDiagnost and OmniDiagnost systems have 2 actuators that are connected to the system stand by 4 shafts. The position of each of these shafts is secured by a locking plate.

During service of two MultiDiagnost-ELEVA systems, it was noticed that a locking plate was missing. In one of these cases, the shaft was still properly positioned after the system had been in use for 9 years. In the other case, involving a system that had been in use for 5 years, the shaft was no longer properly positioned but was still holding the actuator.

Over time, a missing locking plate may result in the shaft of the corresponding connection point becoming improperly positioned. If the shaft works its way totally out of position, it could result in unexpected stand/table tilt movement that cannot be stopped by the user.