Health Products

Synchro 2/ Guidewire (Soft, Standard) (2019-03-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. Synchro 2/ Standard Guidewire
Manufacturer
Stryker Neurovascular
47900 Bayside Parkway
Fremont
94538
California
UNITED STATES
M00326410
More than 10 numbers, contact manufacturer.A. Synchro 2/ Soft Guidewire
Manufacturer
Stryker Neurovascular
47900 Bayside Parkway
Fremont
94538
California
UNITED STATES
M00326010

Introduction

Affected Products
  1. Synchro 2/ Soft Guidewire
  2. Synchro 2/ Standard Guidewire

Reason

Stryker Neurovascular has become aware that the product labels (pouch and carton) for Synchro 2 products contain an incorrect value in millimeters (0.036 mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014 in) on the label is correct. The correct values are 0.014 in (0.36mm). 0.036 mm represents a typographical error.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime