Health Products

WalkAway plus system - 96 - instrument (2019-02-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
39630290WalkAway plus system - 96 - instrument
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B1018-284

Introduction

Affected Products

WalkAway plus system - 96 - instrument

Reason

As a result of an internal investigation for electromagnetic compatibility (EMC) requirements testing, Beckman Coulter has confirmed that WalkAway plus system - 96 - instrument can reset and lose communication when delivered with a 2 kV power surge.