VITROS XT 7600 Integrated System (2019-03-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
J76000142VITROS XT 7600 Integrated System
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
684 4461

Introduction

Affected Products

VITROS XT 7600 Integrated System

Reason

Ortho Clinical Diagnostics identified this issue during internal testing. When the unreported sample dispense error occurs, either a smaller volume of sample or no sample is dispensed on a slide, and the associated condition code is not reported to the operator via the Status console or Condition Review screen. Samples with increased viscosity are more susceptible to these sample dispense errors.

If an unreported sample dispense error occurs, the expected condition code (TE6-47E or TE6-47J) is not reported, and the test result for the sample is not suppressed. Other algorithms that are designed to detect conditions involving smaller than expected volumes due to bubbles, clots or pressure leaks are functioning as expected.