cobas e801 analyzer (2019-03-17) Mar 17, 2019 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type I
Lot or serial number Companies Brand name Model or catalog number All serial numbers. Manufacturer Roche Diagnostics GMBH Sandhoferstrasse 116 Mannheim 68305 GERMANY cobas e801 analyzer 7682913001
Affected Products
cobas e801 analyzer
Potential signal drops on some cobas 800 e801 modules (cat # 07682913001), due to degradation of ProCell II M in the flowpath, caused by bacterial contamination.