Endo Grasph, Endo Clinch II AutoSuture Grasper (2019-03-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
B. Endo Clinch II AutoSuture Grasper174317
P8D1335PRX
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
A. Endo Grasp AutoSuture Grasper173030

Introduction

Affected Products

A. Endo Grasp AutoSuture Grasper
B. Endo Clinch II AutoSuture Grasper

Reason

Medtronic is issuing a voluntary recall following a review of a production records which indicate that the sterilization method used for these lots were not consistent with the labeling and the approved sterilization method. These products are labeled as sterilized with gamma radiation and in some cases, the devices were subjected to re-sterilization with ethylene oxide or a repeat gamma sterilization cycle. While re-sterilization with ethylene oxide does not impact device performance, it is not consistent with the product labeling.