Health Products

Q-Rad System (2019-03-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
RS590LD SN Range RS590LD12K-1002 to RS590LD14N-1005.
Manufacturer
Carestream Health, Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES
Q-Rad SystemMore than 100 numbers, contact manufacturer.

Introduction

Affected Products

Q-Rad System

Reason

Carestream Health has discovered a potential safety problem that could lead to the breakage of the two steel cables that attach the equipment to its ceiling support mechanism. It has been determined that in some instances when the main steel rope breaks, the safety lock may not be triggered and the safety rope is then providing the only support for the lifting column and support arm. This safety rope is not designed for continuous load and on-going movement of the support arm leading to potential serious injury for users and patients, should the safety rope also fail.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime