VAMP Optima, FloTrac sensor VAMP Optima system (2019-03-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
California
UNITED STATES
B. FloTrac Sensor VAMP Optima SystemMHD8VLTW
All lots.
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
California
UNITED STATES
A. VAMP OptimaVOPTIMAL

Introduction

Affected Products
  1. A. VAMP Optima
  2. FloTrac Sensor VAMP Optima System

Reason

The instructions for use are unclear in the steps for blood sampling and the positioning of the stopcock. If the clinician were to incorrectly position the stopcock, the blood sample will be diluted with saline. This has occurred with drawing 2 mL samples for blood glucose readings. The clinician immediately noted upon testing the saline in the blood sample. When this occurs, the clinician will need to obtain another pure blood sample for testing. This may cause a slight delay in treatment.

In addition, some clinicians are putting the reservoir stopcock in the incorrect position when flushing after a blood draw. This can result in not clearing blood completely from the top of the reservoir and subsequent clot forming. Clotting is an inherent risk when any device is introduced into the vascular system. Since line maintenance is a routine part of patient care, line clotting is unlikely to result in an injury. If clotting does occur, the clinician can change out the system with a minimal delay in treatment or monitoring.

There is a potential for distortion in pressure readings if the stopcock at the reservoir is in the incorrect position. This is also true for the lass sample site.