Zeus IE Anesthesia Workstation (2019-03-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.Zeus IE Anesthesia Workstation
Manufacturer
Draegerwerk AG & Co. KGaA
Moislinger Allee 53-55
Lübeck
23542
GERMANY
MK04000

Introduction

Affected Products

Zeus IE Anesthesia Workstation

Reason

Cases have been reported where the gas measurement of Zeus IE Anesthesia workstations became inactive. Draeger analysis revealed that the deactivations were caused by a too strict supervision of the sample gas pumps which are used in the gas measurement modules. As the characteristic of the pump changes with their usage time, the supervision system deactivates the gas measurement and a corresponding alarm message appears. Consequently, the probability of a gas measurement deactivation is correlated with the total usage time of the sample gas pumps. Long-term tests clearly indicate that the gas measurement is reliable for approximately 7 months assuming an average runtime of 84 hours per week. From that point on the probability of a deactivation increases beyond initial estimations.