IMMULITE, IMMULITE 1000 System-Turbo Intact PTH assay (2019-03-17) Mar 17, 2019 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number 255 and higher. Manufacturer Siemens Healthcare Diagnostics Products Limited Glyn Rhonwy Llanberis, Gwynedd LL55 4EL UNITED KINGDOM B. IMMULITE 1000 System-Turbo Intact PTH assay LSKPTZ 255 and higher. Manufacturer Siemens Healthcare Diagnostics Products Limited Glyn Rhonwy Llanberis, Gwynedd LL55 4EL UNITED KINGDOM A. IMMULITE System-Turbo Intact PTH assay LSKPTZ
Affected Products
IMMULITE System-Turbo Intact PTH assay
IMMULITE 1000 System-Turbo Intact PTH assay
Siemens Healthcare Diagnostics has confirmed the potential for failed adjustment slopes (>1.8) for the IMMULITE / IMMULITE 1000 Turbo iPTH (LSKPTZ) assay with Kit lots of 255 and above. Siemens is currently investigating the root cause of the issue.