Introduction

Affected Products
  1. Giraffe Warmer
  2. Panda Warmer

Reason

GE Healthcare has become aware that the bedside panels, and latch areas on the warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle. If an infant impacts a bedside panel with a cracked or broken latch, the panel can disengage and fall open, no longer protecting the infant from falling from the warmer.