VITROS XT 7600 Integrated System (2019-03-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Software Versions 3.4 and 3.4.1.VITROS XT 7600 Integrated System
Manufacturer
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
684 4461

Introduction

Affected Products

VITROS XT 7600 Integrated System

Reason

Due to an undetected mechanical interference, a MicroSlide may not be properly positioned into the Tip Locator. If this occurs, the sample may not be properly dispensed on the slide. Ortho Clinical Diagnostics identified during internal testing that the threshold used in locating the dispense blade position in the Tip Locator is incorrect. If an undetected MicroSlide position error occurs, a condition code is inadvertently not reported by system and the potentially erroneous test result is generated. VITROS XT MicroSlides, MicroWell and MicroTip assays are NOT affected.