Health Products

PROXIMATE ILS Intraluminal Stapler, Endopath ILS Endoscopic Circular Stapler (2019-03-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
ETHICON ENDO-SURGERY, LLC.
475 CALLE C
GUAYNABO
00969
PUERTO RICO
B. Endopath ILS Endoscopic Circular Stapler with low profile anvilECS21A
More than 100 numbers, contact manufacturer.
Manufacturer
ETHICON ENDO-SURGERY, LLC.
475 CALLE C
GUAYNABO
00969
PUERTO RICO
PROXIMATE ILS Intraluminal Stapler with detachable headCDH21A

Introduction

Affected Products
  1. PROXIMATE ILS Intraluminal Stapler with detachable head
  2. Endopath ILS Endoscopic Circular Stapler with low profile anvil

Reason

Through investigation of complaints and returned products, we have confirmed occurrence of uncut washers and malformed staples with our ILS circular staplers, which can compromise staple line integrity. If a problem with the staple line is not adequately addressed or is not recognized, there is a potential risk of postoperative anastomotic leak, gastrointestinal injury, hemorrhage, or hemorrhagic shock.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime