Diode Laser WOLF 445 nm (2019-03-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
650099
Manufacturer
A.R.C. laser GmbH
Bessemerstraße 14
Nürnberg
90411
GERMANY
Diode Laser WOLF 445 nmLS11000-445

Introduction

Affected Products

Diode Laser WOLF 445 nm

Reason

A software bug was detected by the manufacturer's engineering; it may cause higher power output than the one shown on the unit's screen. It is rare and can only happen if the user deliberately uses a shortcut to adjust the power (instead of using the "+" and "-" icons) along with a specific combination of 4 unique laser parameters (which is a again a rare combination). The laser has been in use worldwide for 2 years and no one to date has encountered this.