Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
syngo.via10496180

Introduction

Affected Products

syngo.via

Reason

Siemens is notifying users of a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure using syngo.ct cardiac function or syngo.ct cardiac planning in syngo.via. This risk is due to a software issue found in the TAVI algorithm. During the automatic procedure a smoothing algorithm subsequently adjusts the shape of the centerline. In unfavourable conditions this smoothing algorithm can result in a slight distortion of the centerline and ultimately in a tilting of the measurement plane compared to the previously defined annulus plane. Any measurements of annular parameters taken under the described conditions might be incorrect due to the tilted measurement plane. The magnitude of the deviation depends on the initial shape of the individual centerline related to the patient's anatomy. The stronger the initial bend in the centerline, the stronger the correction level of the smoothing operation.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime