Health Products

bacT/ALERT FN Plus (2019-03-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4051547
Manufacturer
Biomerieux Inc.
100 Rudolphe Street
Durham
27712
North Carolina
UNITED STATES
bacT/ALERT FN Plus410852

Introduction

Affected Products

bacT/ALERT FN Plus

Reason

The issue was discovered during a complaint investigation. An error message was generated by the VIRTUO instrument that incorrectly indicated the bottle would be expiring before the end of the culture bottle test cycle. This error message was displayed on the VIRTUO system at the customer's site. The human readable expiration date on the bottle label was correct and the bottle had not expired. The source of the error message was determined to be an incorrect expiration date within the 2D barcode on the bottle label.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime