Alinity i Processing Module (2019-04-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
AI01053
Manufacturer
Abbott GmbH & Co. KG
Max-Planck-Ring 2
Wiesbaden
65205
GERMANY
Alinity i Processing Module03R65-01

Introduction

Affected Products

Alinity i Processing Module

Reason

There is a potential for loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. A loose cable connection has the potential to heat up and cause damage to the wire and connector. If this occurs, the operator may detect a burning smell. The cable connection to the reagent cooler is located within the processing module and is not intended to be accessed by the operator. Due to the very low flammability rating of the materials used, any heat damage would be localized to the connectors. The cable connections on the reagent cooler have been redesigned to include more robust connectors.