Health Products

Infusion Sets - SS MF Low Sorbing,  Low Sorb CV SS, CV SS, V/NV DP, V/NV CV SS DEHP Free, V/NV CKV SMS DP, V/NV CKV SS DP, Blood Set - NV, SmartSite Needle-Free Valve Port, Filter (2019-04-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
17087341
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
H. Infusion Set - V/NV CKV SS DP2426-0007
17077398
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
F. Infusion Set - V/NV CV SS DEHP Free2420-0007
17077358
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
G. Infusion Set - V/NV CKV SMS DP2420-0500
19025787
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
D. Infusion Set - CV SS11404930
17087336
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
E. Infusion Set - V/NV DP2200-0500
17087410
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
B. Infusion Set - SS MF Low Sorbing10010453
17087329
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
C. Infusion Set - Low Sorb CV SS10015414
18015339
Manufacturer
BD Switzerland SARL
Route de Crassier 17, Business Park Terre Bonne
Eysins
1262
SWITZERLAND
A. Blood Set - NV, SmartSite Needle-Free Valve Port, Filter2477-0000

Introduction

Affected Products
  1. Blood Set - NV, SmartSite Needle-Free Valve Port, Filter
  2. Infusion Set - SS MF Low Sorbing
  3. Infusion Set - Low Sorb CV SS
  4. Infusion Set - CV SS
  5. Infusion Set - V/NV DP
  6. Infusion Set - V/NV CV SS DEHP Free
  7. Infusion Set - V/NV CKV SMS DP
  8. Infusion Set - V/NV CKV SS DP

Reason

Variation in the wall thickness of the administration sets' pumping segment may prevent the upper or lower occluders within the pump module from completely occluding the pump segment after the door is closed and the set's roller clamp is opened. This can result in flow inaccuracies through the pumping cycle process resulting in an over infusion.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime