A. Blood Set - NV, SmartSite Needle-Free Valve Port, Filter
2477-0000
Introduction
Affected Products
Blood Set - NV, SmartSite Needle-Free Valve Port, Filter
Infusion Set - SS MF Low Sorbing
Infusion Set - Low Sorb CV SS
Infusion Set - CV SS
Infusion Set - V/NV DP
Infusion Set - V/NV CV SS DEHP Free
Infusion Set - V/NV CKV SMS DP
Infusion Set - V/NV CKV SS DP
Reason
Variation in the wall thickness of the administration sets' pumping segment may prevent the upper or lower occluders within the pump module from completely occluding the pump segment after the door is closed and the set's roller clamp is opened. This can result in flow inaccuracies through the pumping cycle process resulting in an over infusion.