Artis Q (2019-04-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
Artis Q10848280

Introduction

Affected Products

Artis Q

Reason

Siemens is informing users of Artis systems equipped with a certain motor control unit, about a software error delivered starting May 2018. In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement. Usually activation of a collision sensor blocks all system movement but the floating tabletop, which is still possible to move. However, in the affected Artis systems movement of the floating tabletop is blocked as well.