SOMATOM Force (2019-04-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
75503
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
SOMATOM Force10742326

Introduction

Affected Products

SOMATOM Force

Reason

Siemens is notifying users of SOMATOM Force systems that the current attachment and routing for the W548 cable may result in cable abrasion or breakage. Damage to the MV cable may cause sporadic interference with gantry rotation resulting in scan aborts. While the risk level for scan aborts is below the Siemens risk threshold, troubleshooting this sporatic error is very difficult. Changing the cable attachment will improve the cable routing so the MV cable is no longer subject to possible abrasion or breakage.