Stratus CS 200 Acute Care Diagnostics System (2019-03-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
13050088
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, P.O. Box 6101
Newark
19714
Delaware
UNITED STATES
Stratus CS 200 Acute Care Diagnostics System10714016

Introduction

Affected Products

Stratus CS 200 Acute Care Diagnostics System

Reason

While using a POCT-1A communication protocol, the following error codes will not be sent from the Stratus CS 200 to the LIS.
- CH1H
- CH1L
- CH1M
- CH1N
- CH2H
- CH3H
- CS1K
- CS1V
The error messages will show on the instrument display and printout with the exception of CS1v. CS1V will only be displayed on the instrument as it indicates communication with the lis is not working. The other communication protocols are not affected by this issue.