While using a POCT-1A communication protocol, the following error codes will not be sent from the Stratus CS 200 to the LIS.
- CH1H
- CH1L
- CH1M
- CH1N
- CH2H
- CH3H
- CS1K
- CS1V
The error messages will show on the instrument display and printout with the exception of CS1v. CS1V will only be displayed on the instrument as it indicates communication with the lis is not working. The other communication protocols are not affected by this issue.