Health Products

CARESCAPE Monitors B450, B650, and B850 (2018-09-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Finland OY
Kuortaneenkatu 2,
Helsinki
00510
FINLAND
D. CARESCAPE Monitor B850 System2020913-002
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Finland OY
Kuortaneenkatu 2,
Helsinki
00510
FINLAND
B. CARESCAPE Monitor B450 - Main UnitCARESCAPE Monitor B450
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Finland OY
Kuortaneenkatu 2,
Helsinki
00510
FINLAND
C. CARESCAPE Monitor B650CARESCAPE Monitor B650
More than 10 numbers, contact manufacturer.
Manufacturer
GE Healthcare Finland OY
Kuortaneenkatu 2,
Helsinki
00510
FINLAND
A. CARESCAPE Monitor B850 System - C1 CPU2020913-004

Introduction

Affected Products

A. CARESCAPE Monitor B850 System - C1 CPU
B. CARESCAPE Monitor B450 - Main Unit
C. CARESCAPE Monitor B650
D. CARESCAPE Monitor B850 System

Reason

The following issue is being addressed by GE Healthcare File No. FMI 36127: when multiple CARESCAPE Monitor B450, B650, or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue had been corrected.
When a network overload occurs for a prolonged time, it is also possible that a CARESCAPE Monitor B850 is not rebooting but loses connection with the patient data module (PDM) and a message ""PDM module removed"" is displayed.
Loss of overall monitoring for a prolonged time may lead to a delay in detection of permanent or irreversible impairment or life-threatening changes in the condition of the patient.