Problem 1
some MX4O patient worn monitors may have excessive battery power consumption. For those MX4O's with revision B software and fully charged batteries may last approximately 25% shorter than is specified in the MX4O instructions for use. This issue affects units operating on both rechargeable and disposable batteries.
Problem 2
Low battery technical alerts may not occur as specified in the instructions for use.
For those MX4O's that:
1) operate with disposable AA batteries, and
2) include optional SP02 used in automatic measurement mode.
The technical alerts that may be absent include:
- "TELE BATTERY LOW",
- "NO SPO2T: BATT LOW",
- "REPLACE TELE BATTERY"