ITS4850A IntelliVue Telemetry System - Monitor (2019-03-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not applicable.ITS4850A IntelliVue Telemetry System - Monitor
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
MX40

Introduction

Affected Products

ITS4850A IntelliVue Telemetry System - Monitor

Reason

Problem 1
some MX4O patient worn monitors may have excessive battery power consumption. For those MX4O's with revision B software and fully charged batteries may last approximately 25% shorter than is specified in the MX4O instructions for use. This issue affects units operating on both rechargeable and disposable batteries.
Problem 2
Low battery technical alerts may not occur as specified in the instructions for use.
For those MX4O's that:
1) operate with disposable AA batteries, and
2) include optional SP02 used in automatic measurement mode.
The technical alerts that may be absent include:
- "TELE BATTERY LOW",
- "NO SPO2T: BATT LOW",
- "REPLACE TELE BATTERY"