Health Products

Getinge Washer-Disinfector (2019-04-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
WAA088620
Manufacturer
Getinge Disinfection AB
Ljungadalsgatan 11
Växjö
35115
SWEDEN
Getinge Washer-Disinfector8668

Introduction

Affected Products

Getinge Washer-Disinfector

Reason

On some units of the model Washer-Disinfector 8668 the pump hose can collapse during the washing process. This error is a result from a faulty rubber compound in the process from supplier. If the issue occurs there will be low water pressure to the spray arms which will lead to slow movement of the spray arm and less water flow. This can result in uncleaned goods after finished process.
Get told about the next health product recall.
Get health product recall alerts