Health Products

TactiSys Quartz Equipment (2019-04-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable.
Manufacturer
St. Jude Medical, Inc
One St. Jude Medical Drive
Saint Paul
55117
Minnesota
UNITED STATES
TactiSys Quartz EquipmentPN-004 400

Introduction

Affected Products

TactiSys Quartz Equipment

Reason

Device operating on software v1.7.0 may display intermittent contact force data during procedures. The device log data fills the allocated disk space, which prevents the storage on new log data. In such instances, the contact force information is displayed in purple indicating that the data is not current and may not be accurate. This data will remain purple until communication is re-established or the disk has room to store log files.