Health Products

PRISMAFLEX ST100, ST150, TPE 2000, HF1000 and HF1400 Sets, OXIRIS Set (2019-04-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots expiring between 2020-03-01 and 2021-03-01.
Manufacturer
Gambro Industries
7 Avenue Lionel Terray, B.P. 126
Meyzieu
69883
FRANCE
D. OXIRIS SetOXIRIS
All lots expiring between 2020-03-01 and 2021-02-01.
Manufacturer
Gambro Industries
7 Avenue Lionel Terray, B.P. 126
Meyzieu
69883
FRANCE
B. PRISMAFLEX TPE 2000 SetPRISMAFLEX TPE 2000
All lots expiring between 2020-03-01 and 2022-02-01.
Manufacturer
Gambro Industries
7 Avenue Lionel Terray, B.P. 126
Meyzieu
69883
FRANCE
C. PRISMAFLEX HF1000 and HF1400 SetsPRISMAFLEX HF1000
All lots expiring between 2020-03-01 and 2021-02-01.
Manufacturer
Gambro Industries
7 Avenue Lionel Terray, B.P. 126
Meyzieu
69883
FRANCE
A. PRISMAFLEX ST100 and ST150 SetPRISMAFLEX ST100 SET

Introduction

Affected Products
  1. PRISMAFLEX ST100 and ST150 Set
  2. PRISMAFLEX TPE 2000 Set
  3. PRISMAFLEX HF1000 and HF1400 Sets
  4. OXIRIS Set

Reason

Baxter has received reports of kinked access lines observed during treatment using PRISMAFLEX sets. If the kink prevents blood flow, it causes the PRISMAFLEX control unit to alarm. The issue has been isolated to a subset of lots, based on production dates. A kinked access line has the potential to cause delay in therapy, blood circuit clotting as a result of reduced blood flow, or hemolysis. There have been no reports of serious injury associated with this issue and any are expected to be unlikely.