Health Products

Alaris Pump Module, Model 8100 (2019-04-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
California
UNITED STATES
Alaris Pump Module, Model 81008100

Introduction

Affected Products

Alaris Pump Module, Model 8100

Reason

This recall is an expansion to Recalls RA-64562 and RA-66700 to include bezels manufactured between April 2011 and June 2017. The component of the Alaris pump module mechanical assembly of concern is the bezel assembly, specifically the bezel posts. The affected mechanical assembly may have separation of one or more of the 6 posts that connect the mechanism frame to the bezel assembly, due to weakened plastic.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime