IntelliVue -Information Centre iX - Software, IntelliBridge System (2019-05-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
B. IntelliBridge System866417
All lots.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
A. IntelliVue -Information Centre iX - SoftwareM3290B

Introduction

Affected Products

A. IntelliVue -Information Centre iX - Software
B. IntelliBridge System

Reason

Data from the B. Braun Space LAN or arcomed UniQueDOC infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patient's chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values. Only the HL7 output from the Patient Information Center iX is affected. This problem only happens when the PIC iX or IBS is configured for regional settings in which a comma is used as the decimal separator.

For example, a pump delivering medication at 5.00 cc shows 500.00 cc in the patient's chart or electronic medical record (EMR).