A. IntelliVue -Information Centre iX - Software
B. IntelliBridge System
Reason
Data from the B. Braun Space LAN or arcomed UniQueDOC infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patient's chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values. Only the HL7 output from the Patient Information Center iX is affected. This problem only happens when the PIC iX or IBS is configured for regional settings in which a comma is used as the decimal separator.
For example, a pump delivering medication at 5.00 cc shows 500.00 cc in the patient's chart or electronic medical record (EMR).