Electronic Gas Blender (2019-04-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
48103
Michigan
UNITED STATES
Electronic Gas Blender801188

Introduction

Affected Products

Electronic Gas Blender

Reason

Terumo Cardiovascular Systems (Terumo CVS) is initiating a voluntary field correction of the Electronic Patient Gas System (EPGS) internal flowmeter included with Terumo System 1 devices because the gas flow rate output of the EPGS, which is reported in the Terumo System 1 Central Control Monitor (CCM), may be inaccurate. Through internal investigation, Terumo CVS identified that the cause of the potential inaccuracies is a defect in the EPGS internal flowmeter. Additionally, while the user may notice that the intended gas flow setpoint (slider) and the delivered gas flow reading on the CCM display do not match, there may be no other notification of the fault condition on the CCM.