Health Products

MEDTRACT ST65 (2019-03-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Mortara Instrument Inc.,
7865 North 86th St.,
MEDTRACT ST65TM55HEXRCE

Introduction

Affected Products

MEDTRACT ST65

Reason

An alternate component for an IGBT (Insulated Gate Bipolar Transistor) was added to a Printed Circuit Board Assembly (PCBA) of some treadmills manufactured between 22-dec-2017 and 03-dec-2018. In some instances, the alternate component will overheat, causing the treadmill to no longer function and potentially stop during a stress test. When the failure occurs while performing a test, the treadmill decelerates gradually following normal protocol.