B. Quadrox-i Neonatal Oxygenator with Integrated Hardshell Reservoir
BO-VKMO 11000
More than 10 numbers, contact manufacturer.
Manufacturer
MAQUET CARDIOPULMONARY GMBH
KEHLER STR. 31, RASTATT, GERMANY, 76437
RASTATT
76437
GERMANY
A. Custom Pack with Softline
BO-HQV 48412
Introduction
Affected Products
Custom Pack with Softline
Quadrox-i Neonatal Oxygenator with Integrated Hardshell Reservoir
Reason
Maquet Cardiopulmonary has received complaints in relation to leakage at the QUADROX-i Neonatal blood inlet connector.
Internal investigations have revealed that the leakage only occurs if the enclosed reducing screw adapter 1/4"" x 3/16"" is screwed onto the blood inlet connector of the QUADROX-i Neonatal Oxygenator. The leakage occurs due to a non-conforming blood inlet connector that, in conjunction with the use of the screw reducing adapter 1/4"" x 3/16"", does not result in a tight connectoradapter connection. The combination of the reducing screw adapter 1/4"" x 3/16"" with the blood outlet connector or the use of the blood inlet connector directly to tubing does not exhibit any leakage.