ADVIA Chemistry Creatine Kinase (2019-05-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
465336
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
ADVIA Chemistry Creatine Kinase10729780

Introduction

Affected Products

ADVIA Chemistry Creatine Kinase

Reason

Siemens Healthcare Diagnostics has confirmed increased incidences of system generated (u,u) flags on calibration, quality control and patient results when using the ADVIA Chemistry Creatine Kinase (CK_L) reagent kit lots 465336, 465663, and 468449 on the ADVIA 1800, 2400, and XPT Chemistry Systems.  As per the ADVIA Chemistry Systems operators guide, samples with u or u flags are not to be reported. Siemens has also confirmed for the affected lots that patient results may be falsely decreased or increased across the analytical range even when the u flags are not generated.