MAHURKAR PC Kit, Qplus SE Kits, Acute High Pressure Triple Lumen Catheter Kit Curved and Straight Extensions, Acute Dual Lumen Catheter Kit Pre-Curved, Acute Dual Lumen Catheter Kits (2019-04-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
F. MAHURKAR Acute Dual Lumen Catheter8813-816009
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
G. MAHURKAR Acute Dual Lumen Catheter Kits8813-793009
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
D. MAHURKAR Acute High Pressure Triple Lumen Catheter Kit Straight Extensions8888340629HP
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
E. MAHURKAR Acute Dual Lumen Catheter Kit Pre-Curved8888115132
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
B. MAHURKAR QPlus SE Kits8888135131
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
C. MAHURKAR Acute High Pressure Triple Lumen Catheter Kit Curved Extensions8888345603HP
All lots.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
A. MAHURKAR PC Kit8888115132

Introduction

Affected Products
  1. MAHURKAR PC Kit
  2. MAHURKAR QPlus SE Kits
  3. MAHURKAR Acute High Pressure Triple Lumen Catheter Kit Curved Extensions
  4. MAHURKAR Acute High Pressure Triple Lumen Catheter Kit Straight Extensions
  5. MAHURKAR Acute Dual Lumen Catheter Kit Pre-Curved
  6. MAHURKAR Acute Dual Lumen Catheter
  7. MAHURKAR Acute Dual Lumen Catheter Kits

Reason

In some instances, priming volumes for the MAHURKAR and Argyle acute catheters were established using the maximum calculated volume plus an additional volume to ensure the catheter was fully locked. This approach was chosen to ensure that the lumen is always fully primed, minimizing risk of catheter thrombosis. Therefore, the printed priming volumes on current MAHURKAR And Argyle acute catheters and IFUs exceed the actual locking volume needed to fully fill each lumen. Depending on the catheter size and configuration, a variation from 0.1 mL - 0.5 mL per lumen from the printed priming value (actual locking volume would be lower) could be observed.