SenTec Digital Monitoring System V-Sign Membrane Changer (2019-05-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
SenTec AG
Ringstrasse 39, Therwil,
Basel land
4106
SWITZERLAND
SenTec Digital Monitoring System V-Sign Membrane ChangerMC-I

Introduction

Affected Products

SenTec Digital Monitoring System V-Sign Membrane Changer

Reason

SenTec has initiated a field corrective action to address an issue associated with a manufacturing defect that occurred for certain inserts. Due to a manufacturing issue of the electrolyte cartridge of certain membrane changer inserts, there is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure if a MC with such an insert is used.