Health Products

epoc Blood Analysis System - computing device with custom software (2019-06-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.epoc Blood Analysis System - computing device with custom software
Manufacturer
Epocal Inc.
2060 Walkley Road
Ottawa
K1G 3P5
Ontario
CANADA
EPOC HOST

Introduction

Affected Products

epoc Blood Analysis System - computing device with custom software

Reason

Siemens has confirmed that a "CNC" (could not calculate) result is generated when creatinine is tested on epoc Hosts with system software version 3.30.0 (host software version 3.30.2) in a country language that uses decimal commas instead of decimal points in numbers (e.g. 1,20 compared to 1.20). This issue has been addressed in software version 3.30.1 (host software version 3.30.3).