AXIOM Artis DBA, DTA, BDC magnetic navigation and Artis zeego multi-axis, zee multi-purpose, zee floor, zee ceiling, zee biplane (2019-06-11) Jun 11, 2019 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY H. Artis zee biplane 10094141 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY F. Artis zee floor 10094135 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY G. Artis zee ceiling 10094137 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY D. Artis zeego multi-axis 10280959 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY E. Artis zee multi-purpose 10094139 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY B. AXIOM Artis DTA 7008605 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY C. AXIOM Artis BDC magnetic navigation 59 17 054 Not applicable. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY A. AXIOM Artis DBA 7555357
Affected Products
AXIOM Artis DBA
AXIOM Artis DTA
AXIOM Artis BDC magnetic navigation
Artis zeego multi-axis
Artis zee multi-purpose
Artis zee floor
Artis zee ceiling
Artis zee biplane
Siemens is notifying customers of a potential hardware error on Artis zee and AXIOM Artis Systems with a generator A100. A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.