Health Products

cobas IT middleware (2019-06-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software version: 1.10.00 only
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
cobas IT middleware6498027001

Introduction

Affected Products

cobas IT middleware

Reason

The cobas IT middleware of software version 1.10.00 using the multi-level QC configuration may allow patient results to be released with invalid QC results. This affect the evaluation of the QC lock-status (QC error) when more than one level is configured to be used.