Affected Products
A. ADVIA 120 DIFF TIMEpac
B. ADVIA 2120i DIFF TIMEpac
C. ADVIA 2120 DIFF TIMEpac
Siemens Healthcare Diagnostics has confirmed that platelet clump flags (PLT-CLM) may be generated in the perox channel of the ADVIA 120/2120/2120i Hematology System while running CBC/DIFF (complete blood count / differential) whole blood samples with DIFF TIMEpac reagent lot numbers 94414, 94415 and 31129, in samples without platelet clumps. ADVIA 120/2120/2120i DIFF TIMEpac lot numbers 94414, 94415 and 31129 contain perox 1 reagent lot 94412. The accuracy of the platelet results is not impacted when the perox 1 reagent (lot number 94412) is used on the system and a PLT-CLM flag may be generated due to this issue. Additionally, platelet results are reported from the red blood cell/platelet (RBC/PLT) channel and not measured in the perox channel where platelet clumps are detected. However, the platelet results will be flagged if the clump count is greater than 300, directing the user to investigate the sample.