ADVIA DIFF TIMEpac 120, 2120i, 2120 (2019-06-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
31129
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA 2120i DIFF TIMEpac10312270
94415
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
C. ADVIA 2120 DIFF TIMEpac10312270
94414
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA 120 DIFF TIMEpac10312270

Introduction

Affected Products

A. ADVIA 120 DIFF TIMEpac
B. ADVIA 2120i DIFF TIMEpac
C. ADVIA 2120 DIFF TIMEpac

Reason

Siemens Healthcare Diagnostics has confirmed that platelet clump flags (PLT-CLM) may be generated in the perox channel of the ADVIA 120/2120/2120i Hematology System while running CBC/DIFF (complete blood count / differential) whole blood samples with DIFF TIMEpac reagent lot numbers 94414, 94415 and 31129, in samples without platelet clumps. ADVIA 120/2120/2120i DIFF TIMEpac lot numbers 94414, 94415 and 31129 contain perox 1 reagent lot 94412. The accuracy of the platelet results is not impacted when the perox 1 reagent (lot number 94412) is used on the system and a PLT-CLM flag may be generated due to this issue. Additionally, platelet results are reported from the red blood cell/platelet (RBC/PLT) channel and not measured in the perox channel where platelet clumps are detected. However, the platelet results will be flagged if the clump count is greater than 300, directing the user to investigate the sample.